DAPNIA-01-10 |
---|
320g Ionization-Heat Cryogenic Detector for Dark Matter Search in the EDELWEISS Experiment |
O. Martineau, M. Chapellier, G. Chardin, M. De Jésus, J. Gascon, S. Hervé, A. Juillard, M. Loidl, X-F. Navick, G. Nollez |
Abstract : ======== The EDELWEISS experiment is presently using for direct WIMP detection a 320g heat-and-ionization cryogenic Ge detector operated in a low-background environment in the Laboratoire Souterrain de Modane (LSM). This detector presents an increase of more than 4 times the mass of previous detectors. Calibrations of this detector are used to determine its energy resolution and fiducial volume, and to optimize the detector design for the 1kg phase of the EDELWEISS-I experiment. Analysis of the calibrations and characteristics of a first series of 320g-detectors are presented. |